Abstract
ISO 5359:2008 specifies requirements for low-pressure hose assemblies intended for use with the following medical gases:
- oxygen;
- nitrous oxide;
- medical air;
- helium;
- carbon dioxide;
- xenon;
- specified mixtures of the gases listed above;
- oxygen-enriched air;
- air for driving surgical tools;
- nitrogen for driving surgical tools;
- vacuum.
It is intended in particular to ensure gas-specificity and to prevent cross-connection between systems conveying different gases. These hoses assemblies are intended for use at maximum operating pressures less than 1 400 kPa.
ISO 5359:2008 specifies the allocation of non-interchangeable screw-threaded (NIST) connectors, diameter-index safety system (DISS) connectors and sleeve indexed system (SIS) connectors to medical gases and specifies the dimensions of non-interchangeable screw-threaded (NIST) connectors.
General information
-
Status: WithdrawnPublication date: 2008-06Stage: Withdrawal of International Standard [95.99]
-
Edition: 3Number of pages: 28
-
Technical Committee :ISO/TC 121/SC 1
- RSS updates
Life cycle
-
Previously
WithdrawnISO 5359:2000
-
Now
-
00
Preliminary
-
10
Proposal
-
20
Preparatory
-
30
Committee
-
40
Enquiry
-
50
Approval
-
60
Publication
-
90
Review
-
95
Withdrawal
Amendments
Provide additional content; available for purchase; not included in the text of the existing standard.WithdrawnISO 5359:2008/Amd 1:2011
-
00
-
Revised by
PublishedISO 5359:2014